Urinalysis: diagnostic performance of urine dipstick compared to an automated microscopic method

San Miguel Garay, Erik Alejandro y Llaca Díaz, Jorge M. y Robles Espino, Diana G y Díaz Chuc, Erik A. y Luna Falcón, Claudio Fabrizio y Ayala de la Cruz, Sergio (2023) Urinalysis: diagnostic performance of urine dipstick compared to an automated microscopic method. Medicina universitaria, 24 (4). pp. 177-183. ISSN 1665-5796

[img]
Vista previa
Texto
24680.pdf - Versión Publicada
Available under License Creative Commons Attribution Non-commercial No Derivatives.

Download (244kB) | Vista previa
URL o página oficial: https://www.medicinauniversitaria.org

Resumen

Introduction: Urinalysis is one of the most important clinical laboratory tests because numerous pathologies can manifest or be suspected through this test. Although the previous reports mention that urinary microscopy is a fundamental part of urinalysis for diagnostic support of various conditions, there is a debate about the utility of this test section in a certain patient population. The aim of this study was to determine the diagnostic performance of the urinary dipstick analysis and the potential risks of false-negative (FN) results. Material and methods: This is a retrospective and observational study, and urinalysis information was obtained from non-hospitalized patients. The dipstick and microscopic analyses were performed using the Clinitek-ATLAS (index test) and iQ200-SPRINT (reference standard) devices. Dipstick or microscopy analyses were positive if ≥ 1 parameters were abnormal. A Bayesian hierarchal beta-binomial model was carried out for each performance parameter. Risk analysis was performed as proposed in the literature. Results: Five hundred and fifty-two patients were included in the study. The posterior median at group level was 94% (credible interval 95% [CrI 95%] 89.9-97%) for sensitivity (Se), 57.1% (CrI 95%, 50.1-64.1%) for specificity, and 5.8% (CrI 95%, 2.59-9.64%) for FN rate (FNR). The posterior probability Se > 90% was 95.9% at a group level. The risk analysis found only low-risk false-negative events. Conclusions: The performance of the dipstick analysis was appropriate, with a good certainty of Se > 90% and a FNR < 10% at the operator level. Omission of microscopic analysis can be a safe action in a patient with a negative dipstick since FNs with a clinical impact are not expected.

Tipo de elemento: Article
Palabras claves no controlados: Análisis de orina, Tira reactiva de orina, Microscopía de orina, Análisis de riesgo, Rendimiento de la tira reactiva
Materias: R Medicina > RB Patología
Divisiones: Medicina
Usuario depositante: Dr. Sergio Ayala De la Cruz
Creadores:
CreadorEmailORCID
San Miguel Garay, Erik Alejandrosanmiguel92@gmail.comNO ESPECIFICADO
Llaca Díaz, Jorge M.NO ESPECIFICADONO ESPECIFICADO
Robles Espino, Diana GNO ESPECIFICADONO ESPECIFICADO
Díaz Chuc, Erik A.NO ESPECIFICADONO ESPECIFICADO
Luna Falcón, Claudio Fabriziofabrizioluna_10@hotmail.comNO ESPECIFICADO
Ayala de la Cruz, Sergiosergio.ayalad@uanl.edu.mxorcid.org/0000-0001-5184-3532
Fecha del depósito: 31 Ene 2023 21:48
Última modificación: 31 Ene 2023 21:49
URI: http://eprints.uanl.mx/id/eprint/24680

Versión disponible de este elemento

Actions (login required)

Ver elemento Ver elemento

Downloads

Downloads per month over past year